WASTE ANESTHETIC GAS DISPOSAL

Inspected WAGD exhaust.
Location — away from fresh air intakes
Environments Considerations

• Inspected WAGD source equipment for proper design, installation, and function.
Sizing and Capacity

• Inspected required source and riser valves.
Location/Labeling                                  Valve Handles Locked or Secure

• Tested all area and master alarm functions.

• Inspected proper pressure switch and area alarm sensor locations.

• Inspected minimum of 12 inches Hg of WAGD at source and at each zone.

• Tested each WAGD inlet for maximum available flow rate SCFM.
All inlets shall draw 3.0 scfin without reducing the WAGD pressure below 12 inches Hg at any adjacent station inlet.

• Tested all WAGD zone valves for proper function, identification, and mechanical integrity.

• Tested each WAGD inlet for proper function, identification, adaptation, and mechanical integrity.
Identified leaking and or malfunctioning inlets
Identified inlets delivering substandard vacuum/flow rate

• Checked WAGD system components for proper installation (i.e.: Brazing, Labeling, Hanging, etc.).

• Checked for cross-connections.

 

 
         
 
Health Choice Enterprises, LLC
Phone:800-957-4758
Southwest Region: 505-217-9413
Fax: 801-991-9517
Email: healthchoice@insightbb.com
  Mailing Address
P.O. Box 8112
New Albany, IN 47151
Physical Address:
7090 State Road 311
Sellersburg, IN 47172
         
© Copyright HCE LLC 2009. All Rights reserved. Site design by Skyline Web Solutions